Blumenthal EZ, Frenkel S.

Eye. 2005 Mar;19(3):308-11.
Department of Ophthalmology, Hadassah University Hospital, Jerusalem, Israel. eblumenthal@md.huji.ac.il

ABSTRACT:

OBJECTIVE: To evaluate the interdevice reproducibility of retinal nerve fibre layer (RNFL) thickness measurements obtained with the commercially available GDx-VCC, a scanning laser polarimeter with variable (individualized) corneal compensation.

METHODS: A prospective instrument validation study in which 13 GDx-VCC devices were tested. One eye each, from three normal subjects were used to test each of the devices, on the same day, by an experienced operator. Variability and reproducibility for each of five GDx parameters were calculated.

RESULTS: For each of five tested GDx parameters, the coefficient of variation and 95% confidence interval range (microm), for the 13 devices, respectively, were: TSNIT avg: 5.1%, 3.84 microm; Superior avg: 5.3%, 4.82 microm; Inferior avg: 6.1%, 5.50 microm; TSNIT standard deviation: 8.6%, 2.92 microm; and nerve fibre indicator (NFI): N/A, 5.69. Item reliability (Cronbach’s alpha) for the five GDx parameters are: TSNIT-Avg: 0.97, Sup-Avg: 1.00, Inf-Avg: 0.84, TSNIT-SD: 0.99, NFI: 0.99.

CONCLUSIONS: With the commercially available GDx-VCC, our results indicate that RNFL measurements appear reproducible across devices.

המאמר המלא- כקובץ PDF

 

גלאוקומה וקטרקט:
www.glaucoma.org.il

ניתוח

אבחנה

אתר זה נכתב ע”י פרופ’ איתן בלומנטל. באתר תמצאו אינפורמציה מעמיקה על האבחון והטיפול בגלאוקומה ובקטרקט. האינפורמציה באתר איננה תחליף לייעוץ רפואי.

פוסט קשור

PDF icon 1
Risk Factors and Incidence of Ocular Hypertension After Penetrating Keratoplasty
Faik Oruc¸oglu, MD, Eytan Z. Blumenthal, MD, Joseph Frucht-Pery, MD, and Abraham Solomon, MD Journal of Glaucoma. 2013 Feb 19. Kudret Eye Hospital, Istanbul, Turkey Department of Ophthalmology, Hadassah-Hebrew...
PDF filing cabinet
רשימת פרסומים מדעיים של פרופ' איתן בלומנטל
ORIGINAL RESEARCH PAPERS Blumenthal EZ. Adjusting the slit-lamp oculars: An unnecessary burden or a must?  Survey of Ophthalmology 1995;40(3):225-228. Abstract Full-text (PDF) Blumenthal EZ, Serpetopoulos...
PDF icon 1
Adjusting The Slit-Lamp Oculars: An Unnecessary Burden Or A Must?
Blumenthal EZ. Surv Ophthalmol. 1995 Nov-Dec;40(3):225-8. Department of Opthalmology, Hadassah University Hospital, Jerusalem, Israel. ABSTRACT: Adjusting the slit-lamp eyepiece rings is perceived by some...
PDF icon 1
On Focusing The Slit-Lamp: Part I. An Inaccurate Ocular Setting–What Is There To Lose?
Blumenthal EZ, Serpetopoulos CN. Surv Ophthalmol. 1998 Jan-Feb;42(4):351-4. Department of Ophthalmology, Hadassah University Hospital, Jerusalem, Israel. ABSTRACT: Adjustments of the diopter rings on the...
PDF icon 1
Glaucomatous-Like Cupping Associated With Slow-Growing Supra-Sellar Intracranial Lesions
Blumenthal EZ, Girkin CA, Dotan S. Neuro-Ophthalmology. 2006;30:111-115. Department of Ophthalmology, Hebrew University-Hadassah Medical Center, Jerusalem, Israel. eblumenthal@md.huji.ac.il ABSTRACT: PURPOSE:...