Horani A, Frenkel S, Blumenthal EZ.

Eur J Ophthalmol. 2007 Mar-Apr;17(2):203-7.
Department of Ophthalmology, Hadassah University Hospital, and Hebrew University-Hadassah Medical School, Jerusalem 91120, Israel.

ABSTRACT:

PURPOSE: To quantify the magnitude of test-retest variability (TRV) for normal subjects in serial visual fields (VF) using the frequency doubling technology (FDT) instrument.

METHODS: Twenty-one healthy adults, aged 23 to 60 years, underwent four serial FDT VF tests, using the full-threshold C-20 program of the Zeiss-Humphrey FDT analyzer, on one randomly chosen eye. The VF tests were spaced 2 to 4 days apart. All subjects performed two preliminary FDT tests in order to minimize any learning effect. Test-retest variability was calculated as the standard deviation of each location’s sensitivity value across the four VF tests.

RESULTS: Mean TRV (+/-SD) for the entire field was 2.44+/-1.32 dB. Mean TRV (+/-SD) for the superior, inferior, nasal, and temporal hemifields were 2.48+/-1.3, 2.40+/-1.4, 2.40+/-1.3, and 2.48+/-1.3 dB, respectively. Mean TRV (+/-SD) for the foveal location, the 4 central, and the 12 peripheral locations were 2.49+/-1.4, 2.16+/-1.2, and 2.54+/-1.4 dB, respectively.

CONCLUSIONS: TRV was found to be rather uniform across the visual field of the commercially available FDT device, with only a mild, clinically insignificant, effect of both eccentricity and age on TRV. Variability in the FDT VF, for normal subjects, was found to be more uniform than that of both standard and short wavelength automated perimetry. In addition, a strong inverse correlation was found, in normal subjects, between the mean sensitivity and TRV.

המאמר המלא- כקובץ PDF

 

גלאוקומה וקטרקט:
www.glaucoma.org.il

ניתוח

אבחנה

אתר זה נכתב ע”י פרופ’ איתן בלומנטל. באתר תמצאו אינפורמציה מעמיקה על האבחון והטיפול בגלאוקומה ובקטרקט. האינפורמציה באתר איננה תחליף לייעוץ רפואי.

פוסט קשור

PDF icon 1
Risk Factors and Incidence of Ocular Hypertension After Penetrating Keratoplasty
Faik Oruc¸oglu, MD, Eytan Z. Blumenthal, MD, Joseph Frucht-Pery, MD, and Abraham Solomon, MD Journal of Glaucoma. 2013 Feb 19. Kudret Eye Hospital, Istanbul, Turkey Department of Ophthalmology, Hadassah-Hebrew...
ניתוח עיניים
Katena Blumenthal Conjunctiva Dissector
Dr. Eytan Z. Blumenthal of Jerusalem, Israel, invented and developed this new instrument (Katena K3-1124) for dissection of the trabeculectomy bleb pocket. It features a very smooth round, 2mm wide, disc-shaped...
video-icon-dark-blue
Katena Blumenthal Conjunctiva Dissector- Tutorial CD
[button link=”https://www.glaucoma.co.il/videos/katena-blumenthal-conjunctiva-dissector-basic-background-presentation” type=”big”]Basic background presentation[/button] (6 minutes) [button...
video-icon-turqise
Katena Blumenthal conjunctiva dissector – Basic background presentation (video, 6 minutes)
  [button link=”https://www.glaucoma.co.il/videos/katena-blumenthal-conjuctiva-dissector-routine-technique-video” type=”big”]Routine technique video[/button] (9 minutes) [button...
video-icon-red
Katena Blumenthal conjunctiva dissector: Routine technique (video, 9 minutes)
  [button link=”https://www.glaucoma.co.il/videos/katena-blumenthal-conjunctiva-dissector-basic-background-presentation” type=”big”]Basic background presentation[/button] (6...